Abstract

During residue analysis in complex matrices for food safety purposes, interfering signals can sometimes overlap with those of the analyte of interest. Access to an additional separation dimension besides chromatographic and mass separation, such as ion mobility, can aid in removing interfering signals, allowing for correct analyte identification in these cases. In our laboratory, during routine LC−MS/MS analysis of liver samples for growth promoter residues, an interfering signal was found that matches the retention time and m/z values for stanozolol, a synthetic anabolic steroid. In the present work, the performance of a liquid chromatography coupled to ion mobility mass spectrometry (LC−IM−MS) method has been evaluated to study whether this LC−MS/MS false positive in liver samples could be eliminated by LC−IM−MS analysis. A cyclic ion mobility system already allowed the separation of stanozolol from the interfering peak after only one pass, showing a significant improvement compared to the conventional LC−MS/MS method. Additionally, collisional cross section (CCS) values were calculated and successfully compared with those from literature for identification purposes, eventually allowing both the identification and quantification of stanozolol in this complex matrix.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call