Abstract
We are implementing an X-ray breast Computed Tomography (CT) system on the gantry of a dedicated single photon emission tomography system for breast Tc-99 imaging. For the breast CT system we investigated the relevance of the beam hardening artifact. We studied the use of a single photon counting silicon pixel detector (0.3 mm thick, 256 × 256 pixel, 55 μ m pitch, bump-bonded to the Medipix2 photon counting readout chip) as detector unit in our X-ray CT system. We evaluated the beam hardening “cupping” artifact using homogeneous PMMA slabs and phantoms up to 14 cm in diameter, used as uncompressed breast tissue phantoms, imaged with a tungsten anode tube at 80 kVp with 4.2 mm Al filtration. For beam hardening evaluation we used a bimodal energy model. The CT data show a “cupping” artifact going from 4% (4-cm thick material) to 18% (14-cm thick material). This huge artifacts is influenced by the low detection efficiency and the charge sharing effect of the silicon pixel detector.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.