Abstract
Many conventional ex vivo magnetic resonance imaging (MRI) setups utilize cylindrical or other nonspherical tissue containers which can cause static field (B0 ) inhomogeneity affecting the accuracy of the measurements in an orientation-dependent manner. In this work we demonstrate an experimental method to obtain MRI of ex vivo tissue samples held in a spherical container in order to minimize bulk susceptibility-induced B0 inhomogeneity in arbitrary orientations. B0 inhomogeneity caused by tissue-air susceptibility mismatch can be theoretically eliminated if the surface of susceptibility discontinuity is spherical. This situation can be approximated by putting a tissue sample in a spherical shell filled with materials with tissue-like magnetic susceptibility. We achieved this on an intact monkey brain by (a) holding the brain with a three-dimensional (3D)-printed holder with tissue-like (within 0.5ppm) susceptibility, and (b) enclosing the brain and the holder in an acrylic spherical shell filled with diamagnetic liquid. Furthermore, the sphere and the radio-frequency coil for MRI were mounted on a 3D-printed frame designed to reduce B0 inhomogeneity contributions. The sphere could be rotated freely without disturbing the RF coil to facilitate multi-orientation imaging. We verified our setup by mapping B0 in the monkey brain at 13 different orientations in a human 7T scanner, and measuring orientation-dependent relaxation rates in the white and gray matters of the brain. The results were then compared with a setup where the brain was held inside a cylindrical container. In all orientations, the B0 standard deviation in the brain in the spherical setup (converted to Larmor frequency offset) was less than about 10Hz. This corresponds to two-sigma deviation of B0 of <0.07ppm. The B0 gradient was <9Hz/mm in 95 % of the brain voxels in all orientations. In high-resolution imaging with e.g. voxel size <0.4mm, this corresponds to voxel line broadening of <4Hz (0.013ppm). in the corpus callosum showed distinctly different orientation dependence compared to the gray matter. The B0 uniformity and reliability were much reduced in the cylindrical container setup. We have demonstrated an experimental method to effectively minimize bulk susceptibility-induced B0 perturbation in multi-orientation ex vivo MRI. The method promises to benefit a range of tissue orientation-dependent MR property studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.