Abstract
The measurement of stereotactic radiosurgery dose distributions requires an integrating, high-resolution dosimeter capable of providing a spatial map of absorbed dose. This paper describes the use of a commercial radiochromic dye film (GafChromic MD-55-2) to measure radiosurgery dose distributions with 6 MV X-rays in a head phantom. The response of the MD-55-2 was evaluated by digitizing and analyzing the films with conventional computer systems. Radiosurgery dose distributions were measured using the radiochromic film in a spherical acrylic phantom of 16 cm diameter undergoing a typical SRS treatment as a patient, and were compared with dose distributions provided by the treatment planning system. The comparison lead to mean radial differences of ±0.6 mm, ±0.9 mm, ±1.3 mm, ±1.9 mm, and ±2.8 mm, for the 80, 60, 50, 40, and 30% isodose curves, respectively. It is concluded that the radiochromic film is a convenient and useful tool for radiosurgery treatment planning validation.