Abstract High temperature (570–750°C) three-point impedance measurements on the system Pt/YSZ using a platinum reference electrode exhibited distortions/artifacts due to the voltage divider effect. These distortions could be avoided by using a lower impedance reference electrode (Ag) or employing an instrument with a higher input impedance. A correction procedure was also successful.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.