Abstract
Impedance studies on test circuits demonstrated that three-point measurements are susceptible to voltage divider distortion/artifacts if the reference electrode impedance approaches the analyzer input impedance. These features can be avoided by using an instrument with a high input impedance or a reference electrode with a low input impedance. A correction procedure is also presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.