Abstract

In external beam radiation therapy, it is important to calculate the output of the linear accelerator. The head scatter factor, S h, is one of the important factors for calculation of Monitor Unit, which changes with the size of the irradiation field. In irregular fields shaped by multileaf collimators (MLCs), it is difficult to calculate S h precisely. S h comprises backscatter from the upper and lower secondary collimators and scatter from the flattening filter. We measured the effect of backscatter on a monitor chamber (S b), and we modeled the scatter from a flattening filter using a Gaussian distribution. The modeled parameters used in this method are derived from measurements of square field sizes on the central axis. Furthermore, we divided an MLC irregular field in the shape of fans and integrated the scatter from a flattening filter by a method similar to Clarkson's sector integration. We were able to calculate S h with <1% error in comparison with measurements, even with a field setting with an error of >3% by the conventional method. This method requires no special measuring tools or software.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.