Abstract

In recent years, the use of high-intensity ultrasound equipment has increased in both medical and industrial fields. Consequently, we proposed a high-intensity ultrasound source system using a cylindrical acoustic waveguide for the development of an ultrasound source for the absolute calibration of the receiving sensitivity of hydrophones. We fabricated the experimental ultrasound source system using the cylindrical acoustic waveguide with particular considerations of the material selection and cutting methods to be used. The measured spatial distribution of acoustic pressure generated by our fabricated ultrasound source almost coincides with the results obtained by numerical simulation. In particular, the measured and simulated −6 dB main beam widths agreed within 15%. The linearity was measured using ultrasound waveforms with applied peak-to-peak voltages within the linear range of the power amplifier. Nonlinear phenomena were observed for the measured acoustic pressures of about 130 kPa or more.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.