Abstract
In this paper we present description of the developed current DLTS spectrometer to investigate deep energy levels in barrier structures based on crystalline and non-crystalline semiconductors with large rectifying junction area and high leakage currents. This modified equipment allows to define DL parameters such as ionization energy, cross section and concentration in the wide temperature range 7…500К, relaxation time range 10 μs…10 s, maximum barrier capacitance till 2000 pF, relaxation current amplitude range 10 pА…10 mА.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.