Abstract
A procedure based on solid-phase microextraction (SPME) and gas chromatography coupled with mass spectrometry (GC-MS) was developed and validated in order to analyse 10 phenols in water samples. The optimised conditions were obtained using polyacrylate fibre (PA), 20ml of sample volume, 10% NaCl, pH 4.0 and direct extraction at 35 degrees C and 1000rpm, for 40min. The linear range and quantification limits for these compounds by SPME-GC-MS were defined. An evaluation of the main uncertainty sources of this method is included, which allows expanded uncertainties in the 9.4-35% range for the majority of the compounds. The main source of uncertainty is associated with matrix effects. The validated method is suitable for monitoring the production and distribution of potable water and was used, in field trials, for the analysis of samples from main intakes of water (surface or underground) and from water supply system of a large area (Lisbon and neighbour municipalities).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.