Abstract

The recently released IEC 61066:2006 standard [IEC 61066:2006. International Standard, Thermoluminescence Dosimetry Systems for Personal and Environmental Monitoring] is a new guideline for type testing a TLD dosimetry system. A few main changes made in this standard include harmonizing the reference radiation and calibration with ISO standards, integrating the basic uncertainty analysis, and aligning IEC uncertainty requirements on dosimetry systems with those stated in ICRP Publication 75. We tested to this standard with an advanced Harshaw TLD system: Model 8800 Plus Reader and LiF:Mg,Cu,P dosemeter. This work describes the more than 10 tests in four major categories and provides the learning experience in uncertainty analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.