Abstract
Histotripsy is a non-invasive high-intensity focused ultrasound (HIFU) based therapy using the mechanical effect of cavitation bubbles. In this paper, we experimentally examined the effectiveness of a proposed method of scanning the HIFU focus in the direction of propagation to expand the region of the bubble cloud and increase the treatment throughput. First, the continuous generation of cavitation bubbles was observed with a high-speed camera, and it was confirmed that the region of bubble cloud generation could be expanded by focus scanning. Next, we performed ex vivo experiments. Tissue treated with histotripsy is disrupted and homogenized at the subcellular level, resulting in a decrease in the backscatter intensity of the ultrasound. Therefore, we observed the progression of tissue homogenization with histotripsy by evaluating the reduction of echo intensity in ultrasound imaging. The results suggest that focus scanning can improve the time efficiency of treatment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.