Abstract
We aimed to provide an unbiased estimate of the confounding effect of subcutaneous fat thickness on ultrasound echo intensity (EI) measures of muscle quality. The effect of fat thickness on EI was verified for an approximate range of 0 to 3 cm of fat using exogeneous layers of pork fat over the human tibialis anterior muscle. Sonograms were obtained (i) with focus constant across fat thickness conditions, and (ii) with focus position adjusted to the muscle region of interest (ROI) position for each fat thickness level. In agreement with our hypothesis, increasing fat between the probe and the ROI resulted in a decrease in EI. This overestimating effect of fat on muscle quality differs between sonograms with constant focus and sonograms with focus position adjusted to the vertical displacement in ROI position that occurs for different levels of fat thickness. Correcting equations to account for the overestimating effect of fat on muscle quality are provided for both focus conditions. This is the first study to systematically analyze the confounding effect of fat thickness as an independent factor and the provided equations can be used for improved accuracy in estimates of muscle quality in obese/overweight subjects/patients. Novelty: The independent confounding effect of subcutaneous fat thickness on ultrasound (US) estimates of muscle quality was quantified. US estimates of muscle quality depend on whether focus is adjusted to the muscle region of interest or not. Equations for correcting muscle quality estimates are provided.