Abstract
Pig kidneys (n = 7) were connected to a NMP circuit. PET/MRI was performed at two different pump flow levels: a blood-oxygenation-level-dependent (BOLD) MRI sequence performed simultaneously with a H215O PET sequence for determination of RBP. RBP was measured using H215O PET in all kidneys (flow 1: 0.42-0.76 mL/min/g, flow 2: 0.7-1.6 mL/min/g). We found a linear correlation between changes in delivered blood flow from the perfusion pump and changes in the measured RBP using PET imaging (r2 = 0.87). Our study demonstrated the feasibility of combined H215O PET/MRI during NMP of isolated porcine kidneys with tissue oxygenation being stable over time. The introduction of H215O PET/MRI in nephrological research could be highly relevant for future pre-transplant kidney evaluation and as a tool for studying renal physiology in healthy and diseased kidneys.