Abstract
A 7 Tesla Magnetic Resonance Imaging (MRI) system with parallel transmission (pTx) for 32 near-magnet Cartesian feedback loop power amplifiers (PA) with output power of 1kW is under construction at Erwin L. Hahn Institute for Magnetic Resonance Imaging. Variation of load impedance due to mutual coupling of neighborhood coils in the array may lead to instability of the Cartesian feedback loop amplifier. MRI safety requires unconditional stability of the PAs at any load. In order to avoid instability in the pTx system, conditions and limits of stability have to be investigated for every possible excitation mode for the coil array. In this work, an efficient method of stability check for an array of two transmit channels (Tx) with Cartesian feedback loop amplifier and a selective excitation mode for the coil array is proposed which allows extension of stability investigations to a large pTx array with any arbitrary excitation mode for the coil array.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.