Abstract

PurposePulmonary vein isolation (PVI) is a cornerstone therapy in patients with symptomatic atrial fibrillation. One current method is performing a PVI using a cryoballoon (CB). The CB is inserted into the left atrium via a steerable sheath. However, at times, passing of the interatrial septum by the sheath is hindered, e.g., due to septal fibrosis. Here we report our experience with an evasion maneuver to facilitate this approach using a 6F multipolar and steerable coronary Sinus catheter (CS) for predilatation of the interatrial septum. Methods and resultsWe report 10 patients undergoing a CB-PVI, where the investigator experienced difficulties in passing the interatrial septum with the CB sheath. In these cases, after three conventional abortive attempts, we predilated the transseptal puncture site using both the CS catheter and the dilatator of the CB sheath. Thereafter access of the CB sheath to the left atrium could be achieved instantly and without further resistance. ConclusionWe report a safe and feasible maneuver to facilitate transseptal access with the CB steerable sheath in cases complicated by excessive interatrial resistance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.