Abstract

The performance of eight high-purity silica-based reversed-phase columns was analysed using chemometric pattern recognition. Tests were performed at pH 3.0 and 7.0, using 10 compounds, three mobile phase modifiers (methanol, tetrahydofuran and acetonitrile) and four parameters {retention factor (k′), column efficiency (N), the Dorsey–Foley efficiency [N(df)] and the asymmetry factor (As])}. Principal components analysis was performed on subsets of the data. Features of both scores and loadings plots, together with the corresponding correlation coefficients between chromatographic tests, are discussed in detail. It is concluded that methanol and acetonitrile exhibit similar properties, the relative performance at each pH differs, the number of compounds used in the test can be reduced to a recommended five, and several of the original tests are redundant. Procedures are recommended for determining a range of tests that provide information on the column performance, which can be applied to other situations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.