Abstract
To examine the torque moment that occurs between esthetic brackets and bendable alloy (stainless steel [SS], titanium-molybdenum [Ti-Mo], and titanium-niobium [Ti-Nb]) wires. This study examined ceramic (CR), zirconium oxide (ZC), polycarbonate (PC), and conventional metallic brackets (MT) (upper, 0.018-inch and 0.022-inch slots) combined with SS, Ti-Mo, and Ti-Nb wires using elastic module ligation. The torque moments delivered by various wire and bracket combinations were measured using a torque gauge apparatus. The wire torque angles at 5-40° were examined. The torque value increased in the order of CR, ZC, MT, and PC brackets for both 0.018-inch and 0.022-inch slots. The fracture points of the CR and ZC brackets combined with SS and Ti-Mo wires were approximately more than 30° and 35°, respectively. No fracture points were detected in the combination of ZC brackets and Ti-Nb wires. The current study identified the material characteristics of CR, ZR, and PC brackets during torque tooth movements. The present results demonstrate a characteristic combined effect between different esthetic brackets and bendable alloy wires.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.