Abstract

(amikacin MIC by Vitek 2 AST-N055 card: ≤2μg/mL, 11 isolates; 4μg/mL, 19 isolates; 8μg/ mL, 17 isolates; 16μg/mL, 27 isolates; and ≥64μg/ mL, 2 isolates) from a university hospital and eval-uated the Vitek 2 AST-N055 card vs the broth micro-dilution as a reference method for testing suscepti-bility to amikacin. Vitek 2 AST-N055 card yielded very major errors in 15 isolates (19.7%) and minor errors in 26 isolates (34.2%). Of the 15 isolates shown very major errors, 14 had Vitek 2 MICs ranging from 8 to 16μg/mL. The results of our study suggest strongly that it is unreliable to test the amikacin sus-ceptibility by Vitek 2 AST-N055 card of

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.