Abstract
Exposure to ultrasound combined with intravenous injection of microbubbles is a technique that can be used to temporarily disrupt the blood–brain barrier. Transcranial monitoring of cavitation can be done with one or more passive cavitation detectors (PCDs). However, the positioning of the PCDs relative to the cavitation site and the attenuation of these signals by the skull are two sources of error in the quantification of cavitation activity. The aim of this study was to evaluate in vitro the amplitude variation of cavitation signals that can be expected for an excised porcine skull model. The variation caused by the relative positioning of the PCD with respect to the cavitation site was quantified. A position-based correction of the signal amplitude was evaluated. Pig skull samples were used to assess variation in signal amplitude caused by bone. The overall coefficient of variation of the signals owing to these measurement biases was estimated at 30.8%.