The Rapid Strep system (API System S.A., Montalieu-Vercieu, France) was evaluated, without additional tests, in the identification of 209 streptococci. Organisms included 59 beta-hemolytic, 36 group D, 24 Streptococcus pneumoniae, and 90 viridans group streptococci. The Rapid Strep system correctly identified to species level 69.5% of the beta-hemolytic strains, 100% of the group D strains, none of the S. pneumoniae strains, and 84.5% of the viridans group streptococci. The method provided excellent identification rates of groups A, B, and D but failed to differentiate between groups C and G. The method for preparation of suspensions for the Rapid Strep system, as initially recommended by the manufacturer, was responsible for the failure to identify S. pneumoniae. This method was subsequently modified to yield a heavier inoculum; all 10 pneumococcal strains tested with the revised inoculum method were correctly identified. Good identification rates of commonly encountered viridans strains were found. The Rapid Strep system represents a worthwhile advance in streptococcal species identification, especially for group D and viridans strains.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.