Abstract

In this study, our aim is to assess the potential of Fricke and polyacrylamide gel (PAG) dosimeters to quantitatively evaluate the efficiency of potential radiation sensitizers/protectors and antioxidants. These compounds are of importance in radiotherapy as well as in disease prevention and promotion of health. The basic principle of the Fricke dosimeter is the radiation-induced oxidation of Fe2+ to Fe3+ in an aerated aqueous 0.4 M H2SO4. The production of ferric ions is most sensitive to the radical species produced in the radiolysis of water. Using this method, we observed that cystamine (one of the best of the known radioprotectors) can prevent oxydation of Fe2+ from reactive radiolysis species. However, one obvious disadvantage of the Fricke dosimeter is that it operates under highly acidic conditions (pH 0.46), which may degrade biological compounds. In contrast, the pH of the polyacrylamide gel (PAG) dosimeter is almost neutral, such that degradation of compounds is less probable. A change in R2-dose sensitivity was observed in the presence of radiosensitizers/radioprotectors and antioxidants. The protective effect of Trolox (a well-known antioxidant) and thiourea (a radioprotector) was readily observed using the PAG dosimeter. Incorporation of iodinated radiation sensitizers such as NaI and an iodine contrast agent led to a quantifiable sensitizer enhancement ratio. These studies suggest that the Fricke and the PAG dosimeters have the potential to evaluate the efficiency of radiation sensitizers/protectors and antioxidants.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call