Abstract

The main objective of this study was the standardization of the direct anti-A, anti-B haemagglutination assay for immunoglobulin products in microtitre plates and gelcards by automation on a liquid handling robot. In addition, the evaluation of the pipetted microtitre plates with a computer-controlled camera was investigated and these results were related to titres from visual live reading.The titres obtained with the automated and the manual assay in microtitre plates and gelcards were compared. They were in excellent agreement: the titres of samples processed with the automated method varied maximally one titre step relative to the median titres of the same sample analysed with the manual method. The implementation of a camera guided test plate reading further increased the repeatability of the reported titres.In summary, the automated haemagglutination assay combined with a camera improved the consistency plus the traceability of the results and reduced the required hands-on time.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.