Abstract
The frequency response of a free running RF generator was evaluated for its ability to provide plasma diagnostic information. A small probe coil in the plasma torch housing and a frequency counter were used for frequency measurement. A linear dependence between the RF generator frequency and the flow of a foreign gas (CO2, H2, He, N2—introduced into the central channel of the ICP) was encountered. Moreover, the inductively coupled plasma (ICP) was found to extinguish at a constant, characteristic frequency when the flow of foreign gas was too high, allowing a simple prediction of the maximal tolerable foreign gas flow. Spectroscopic plasma diagnostic techniques (Mg II/Mg I ratio, excitation temperature and rotational temperature) were also investigated but provided little indication of instable plasma conditions due to the addition of foreign gases.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.