Abstract
An extraction discharge was formed at the tail flame of an Ar inductively coupled plasma (ICP) for atomic emission spectrometry. Analytical performance was measured for 19 elements at 46 different wavelengths for the supplementary discharge and for the Ar ICP alone. Detection limits of the Ar ICP used with the extraction discharge were generally lower than those for the Ar ICP alone. While detection limits measured at Ca II 370.60, Ca II 373.69, Zn II 202.55, Zn I 213.86, Mg II 279.55 Mg II 280.27, Mn II 293.30, Mn II 293.93 and Mn II 294.92 nm were improved by a factor of 8–44 in the presence of the extraction discharge, the enhancement in detection power was not significant (less than a factor of five) for the majority of the elements tested. Compared with the conventional Ar ICP, higher excitation and rotational temperatures were measured for the extraction discharge. Electron number densities measured in the presence of the extraction discharge (at 26 mm) were approximately the same as that of an Ar ICP alone (at 18 mm). These results were compared with the data obtained previously for an extraction discharge and with those found in ICP mass spectrometry.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.