Abstract The electronic response time of the Forward Scattering Spectrometer Probe (FSSP) has been evaluated and is shown to affect the measurement of droplet size. The size of droplets are underestimated when airspeeds exceed 50 m s−1 and size distributions are broadened as a result of nonuniformities in the laser beam of the probe. A processing scheme that requires a detailed knowledge of a probe's optical and electronic characteristics is described that compensates for the measurement errors caused by the probe's electronic and optical limitations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.