Abstract

The objective of this study was to evaluate the performance of the CELL-DYN® 3500 for rat and mouse blood analysis in a routine environment. The WBC (white blood cells), RBC (red blood cells), PLT (platelets) counts and the WBC differential were determined. In addition, the following aspects were studied: within-run precision, day-to-day precision, biasfree paired difference precision; extended ranges of linearity for RBC, HCT (haematocrit), WBC, PLT; carry-over, the fffect of blood ageing, cell stability with different anticoagulants; and the normal ranges, the out of range flagging and some typical pathology cases.