Abstract

Objectives: NGAL (Neutrophil Gelatinase-Associated Lipocalin) has emerged as a new biomarker for the identification of acute kidney injury. Reliable clinical evaluations require a simple, robust test method for NGAL, and knowledge of specimen handling and specimen stability characteristics. We evaluated the performance of a new urine NGAL assay on the ARCHITECT analyzer. Methods: Assay performance characteristics were evaluated using standard protocols. Urine specimen storage requirements were determined and biological variability was assessed in a self-declared apparently healthy population. Results: Assay performance data showed good precision, sensitivity and lot-to-lot reproducibility. There was good short term 2–8 °C sample stability, however, long term storage samples must be kept at − 70 °C or colder. The largest variance component in a biological variance study was within-day. Conclusions: The ARCHITECT NGAL assay proved to be a precise and reproducible assay for the determination of urine NGAL.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.