Abstract
The purpose of this study was to determine if there was agreement between a new point-of-care (POC) lactate analyzer and a handheld laboratory analyzer when measuring blood lactate concentration in cats. Prospective observational study. University veterinary teaching hospital. Fifty-four cats that presented to an emergency service. None. Lactate concentrations as measured by the handheld laboratory analyzer ranged from 0.3 to 15.4 mmol/L. Agreement analysis of the handheld laboratory analyzer and the POC lactate meter demonstrated a bias, -0.06 and limits of agreement ranging from -0.87 to 0.99 mmol/L. Regression analysis demonstrated a coefficient of determination (R2 ) of 0.98. Results of the present study indicate that the POC lactate meter provided results that are in agreement with a handheld laboratory analyzer when measuring lactate in clinically ill patients.