Abstract

To evaluate of the accuracy of domestic commercial HBV DNA real-time polymerase chain reaction kits. Using COBAS TaqMan HBV Test as reference, we evaluate the accuracy of a domestic commercial HBV DNA real-time polymerase chain reaction kit (PG). Among the samples with viral load at the range of 10(1), 10(2), 10(3), 10(4), 10(5), 10(6), 10(7) (IU/mL), the Coefficient of Correlation(r) between the result determined by domestic kit (PG) and those of Roche COBAS TaqMan HBV Test were: -0.08011, -0.05056, 0.105642, 0.312181, 0.908046, 0.866175, -0.23295, respectively; the percentage of false negative results were 60%, 30%, 33.3%, 8.3%, 0, 0, 0, respectively. Among the samples with viral load over than 10(7) (IU/ml), the result determined by PG is significantly lower. The accuracy of PG is not satisfied, especially in those samples with viral load less than 10(4) (IU/ml). A implication from these observation is that samples from patients received antiviral treatment should be tested by Roche COBAS TaqMan HBV Test.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.