Abstract

To evaluate the laser power required to rupture both the retinal vein and Bruch's membrane in a porcine model using a new laser system for treatment of retinal vein occlusion. The retina was treated in areas with and without an overlying vein. Sections through the laser sites were stained for light and transmission electron microscopy. The percentage of the laser sites with rupture of the retinal vein and Bruch's membrane, the degree of tissue damage, and the diameter of the rupture in Bruch's membrane were determined. The rate of vein rupture was 75% at 3 W, and rupture of Bruch's membrane was achieved at all powers tested. The mean diameter of the rupture in Bruch's membrane increased with laser power. The Integre Duo laser system (Ellex Medical Lasers, Adelaide, Australia) is capable of rupturing the retinal vein and Bruch's membrane at lower powers than prior laser systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.