Abstract
Signals of a light-sharing positron emission tomography (PET) detector are commonly multiplexed to three analog pulses (E, X, and Y) and then digitally sampled. From this procedure, the signal energy that are critical to detector performance are obtained. In this paper, different signal energy calculation strategies for a self-developed SiPM-based PET detector, including pulse height and different integration methods, are evaluated in terms of energy resolution and spread of the crystal response in the flood histogram using a root-mean-squared (RMS) index. Results show that integrations outperform the pulse height. Integration using the maximum derivative value of the pulse E as the landmark point and 28 integrated points (448ns) has the best performance in these evaluated methods for our detector. Detector performance in terms of energy and position is improved with this integration method. The proposed methodology is expected to be applicable for other light-sharing PET detectors.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.