Abstract

The sensing modified electrode was prepared using glucose oxidase immobilized onto vanadium pentoxide xerogel with glass/FTO as support electrode to evaluate the possibility to construct a V2O5/GOx Extended Gate Field Effect Transistor biosensor. Previously, our studies exhibited a sensitivity of V2O5 of 58.1 mV/pH. The use of Nafion® onto V2O5/GOx caused a decrease of mass loss after several cycles compared to the modified electrode without Nafion® during the EQCM and cyclic voltammetrics studies. Electrical characterization of V2O5/GOx demonstrated a tendency to stability after 200 s as a function of applied current. In presence of glucose and in different pH, the current decreased when the glucose concentration increased due to the lower active sites of enzyme. After ten voltammetric cycles, the total charge tends to structural stability. In pH = 5.0, the modified electrode based on V2O5/GOx Extended Gate Field Effect Transistor presented more tendency to sensitivity in different concentration of glucose.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.