Abstract

Abstract Introduction/Objective Our cardiac intensive care unit (CICU) asked the laboratory to provide plasma hemoglobin testing to monitor effects of ventricular-assist devices (VAD). Roche hemolysis index (HI) has been reported to be suitable for this purpose. Use of HI as a reportable test, however, must be validated in-house as a laboratory- developed test (LDT), according to CLIA regulations. We describe results of our evaluation, and caveats for intended use. Methods/Case Report Concentrated hemolysate was produced by osmotic lysis of packed red blood cells in water. Hemolysis in plasma was by dilution of concentrated hemolysate. Verification of HI as a measure of hemoglobin (mg/dL) used the Sigma-Aldrich spectrophotometric hemoglobin assay as a gold standard. Linearity, linear range, sensitivity, were investigated by dilution measurements; interferences by admixture experiments; reproducibility (precision, intra- and inter-assay) in each case by replicate measurements (n = 20). Results (if a Case Study enter NA) HI was confirmed to correspond to hemoglobin in mg/dL. Linearity was between 2-1000 mg/dL (r2 >0.99). Intra-assay precisions were <=2.5% (hemoglobin = 74,148 mg/dL); inter-assay precisions were <=4.9% (hemoglobin = 69,139 mg/dL). HI variability was ±2 at very low values (0-10 mg/dL). There were no interferences observed among common therapeutic drugs. Hyperlipemia (evaluated up to lipemic index (LI) = 225) and hyperbilirubinemia (evaluated up to icteric index (II) = 7) demonstrated no significant interference. Conclusion Analytically, HI exhibited acceptable performance for reporting of plasma hemoglobin. Deployment would require establishment of quality control (QC) and proficiency testing procedures for HI. Clinically, a caveat for use is that HI cannot distinguish between hemolysis in circulation vs. that due to sample collection. For the CICU, HI >10 mg/dL is characteristic of more than 20% of samples drawn in general, irrespective of use of VAD. Temporal patterns of HI must therefore be carefully evaluated in parallel with haptoglobin measurements to assist in VAD management.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call