Abstract

Adaptation of the Randox Enzymatic Manual UV Ammonia method to be used on the Roche Cobas 6000 (c501) automated analyzer platform. The Randox ammonia reagent was evaluated for precision, linearity, accuracy and interference from hemolysis, icterus and lipemia on the Roche c501 analyzer. Comparison studies were conducted for the Randox reagent between Roche c501, Siemens Vista, Ortho Vitros 250, and Beckman DxC methods. The Randox reagent demonstrates acceptable within-run (L1=65 μmol/L, CV 3.4% L2=168 μmol/L, CV 1.9%) and between-run precision (L1=29 μmol/L, CV 7.3% L2=102 μmol/L, CV 3.0%), Analytical Measurement Range (7-940 μmol/L), and accuracy. The method interference profile is superior for the Randox method (hemolysis index up to 600, icteric index up to 60, lipemic index up to 100) as compared to the Roche method (hemolysis index up to 200, icteric index up to 10, lipemic index up to 50). Comparison was very good between the Randox reagent and two other wet chemistry platforms. The Randox Enzymatic Manual UV Ammonia reagent is an available alternative to the Roche Cobas c501 reagent. The method is more robust to endogenous interferences and less prone to instrument error flags, thus allowing the majority of clinical specimens to be reported without additional sample handling at our institution.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.