Abstract

Polyurethane foam (PUF) passive air samplers (PAS) are a common and highly useful method of sampling persistent organic pollutants (POP) concentrations in air. PAS calibration is necessary to obtain reasonable and comparable semi-quantitative measures of air concentrations. Various methods are found in the literature concerning PAS calibration. 35 studies on PAS use and calibration are examined here, in conjunction with a study involving 10 PAS deployed concurrently in outdoor air with a low-volume air sampler in order to measure the sampling rates of PUF-PAS for polychlorinated biphenyls (PCBs), polybrominated diphenyl ethers (PBDEs), polycyclic musks (PCMs), and polycyclic aromatic hydrocarbons (PAHs). Based on this analysis it is recommended that (1) PAS should be assumed to represent bulk rather than gas-phase compound concentrations due to the sampling of particle-bound compounds, (2) calibration of PAS sampling rates is more accurately achieved using an active low-volume air sampler rather than depuration compounds since the former measures gas- and particle-phase compounds and does so continuously over the deployment period of the PAS, and (3) homolog-specific sampling rates based on KOA groupings be used in preference to compound/congener-specific or single sampling rates.