Abstract

Abstract Fatty acid methyl esters, 2,6-dimethylaniline, and 2,6-xylenol were used for rapid evaluation of the usefulness of packed GLC columns for analysis of unknowns at the nanomolar level. Three types of column construction with three liquid phases of widely varying polarity were tested by the above procedure. A new type of GLC column, a dynamically coated packed column prepared with a surface active agent, was found to yield better overall performance for less polar liquid phases than packed silanized glass or packed stainless steel columns.