Abstract

AbstractThe polarity of carbon layer open tubular (CLOT) columns coated with a layer of non‐porous graphitized carbon black (Carbopack B) modified with an appropriate amount of polar polyglycol liquid phase has been evaluated and compared with that of standard polar (Supelcowax‐10) and non‐polar (SPB‐1) bonded phase open tubular columns. The efficiency and selectivity were measured at various temperatures and the polarity of the columns was evaluated by use of McReynolds' constants and the difference in apparent carbon number, ΔC of linear alkanes and alcohols. The polarity of the CLOT column was found to depend on temperature, and changing the analytical conditions therefore enabled the separation of compounds of different polarity whose reciprocal position and resolution were affected by temperature. The application of calculation methods which enable programmed temperature retention times to be predicted from isothermal data was also found to be possible when the polarity of the CLOT column changes with temperature.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.