Abstract

In environmental and biological monitoring for carbon-14 and tritium, the presence of color and chemical quenching agents in the samples can degrade the efficiency of liquid scintillation counting. A series of experiments was performed to evaluate the usefulness, under routine conditions, of first oxidizing the samples to improve the counting by removing the color and quenching agents. The scintillation counter was calibrated for the effects of quenching agents on its counting efficiency. Oxidizing apparatus was tested for its ability to accurately recover the /sup 14/C and /sup 3/H in the samples. Scintillation counting efficiences were compared for a variety of oxidized and unoxidized environmental and bioassay samples. The overall conclusion was that, for routine counting, oxidation of such samples is advantageous when they are highly quenched or in solid form.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.