Abstract
Introduction: Ruthenium-106 brachytherapy is a primary treatment for Uveal Melanoma (UM), the most common intraocular malignancy in adults. This study evaluated the safety of Ru-106 applicators at 3 Tesla (T) MRI and their impact on image quality. Methods: Magnetic attraction and eddy currents were tested on a 20mm diameter Ru-106 applicator using a nylon string and a porcine eye. Safety criteria were defined by ocular oncologists, comparing magnetic field interactions to the forces exerted on the eye during surgery. Five UM patients were scanned at 3T MRI with the applicator in-situ using both conventional anatomical sequences and scans optimized to reduce metal artifacts. Results: Minimal magnetic interactions was observed. Eddy currents caused slight lagging during fast movements, and temporary detachment of the applicator of the porcine eye in conditions that were considered unrealistic for clinical scans. Significant susceptibility artifacts compromised image quality of the affected eye. Conclusion: Patients with Ru-106 applicators can be safely used in 3T MRI with some simple precautions. MR-image quality of the eye was poor due to major susceptibility artefacts, however, imaging of extra-ocular anatomy is feasible.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.