Abstract

Electrometers are important devices that are part of the standard dosimetry system. Therefore, we evaluated the variation of electrometer calibration coefficients (kelec) over 1 year in this study. We investigated two types of electrometers: a rate mode and an integrate mode. Each electrometer was connected to a charge generator, a constant charge was applied, and kelec was determined by measuring the current. The current measurements were repeated once a month. For electrometers with multiple ranges, measurements were taken at low and medium ranges. Almost all kelec measurements agreed within 0.2% of the initial measurements. However, the low range of the electrometer with an integrate mode showed seasonal variation, with a variation greater than 0.2%. This study shows that electrometers may exhibit errors that cannot be detected through annual inspections. The importance of quality assurance using a charge generator at one’s own institution was demonstrated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.