A high concentrated glass bead, with a flux to sample ratio of two to one, was prepared to evaluate the potential for major and trace element analyses of various silicate rocks by X-ray fluorescence spectrometry. Careful instrument set up, matrix effect and peak overlap corrections allowed enhanced accuracy of analysis. The accuracy reached a level comparable to that of using five to one glass beads for major elements and pressed powder pellets for trace elements. Using this method, major and trace element analyses were possible using one single glass bead.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.