Abstract

AbstractIn phase‐shifting electron holography, Fresnel diffraction at an electron biprism introduces distortions to the amplitude and phase of the electron waves, which cause phase errors in the reconstructed phase images. To reduce the phase errors, two correction methods have been proposed, termed Fresnel corrections. In order to evaluate the phase‐shifting electron holography with Fresnel corrections, we faithfully simulated holograms by considering the influence of Fresnel diffraction, the quantum noise and the electron wave coherency, and then compared the phase images reconstructed from corrected and uncorrected holograms. We found that the phase error due to Fresnel diffraction was reduced to 1/20 by using Fresnel corrections, and that the phase measurement precision reached 2π/410 rad. We are able to observe the weak electric field due to positive charges corresponding to five electrons. Copyright © 2003 John Wiley & Sons, Ltd.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.