Abstract

Ion mobility mass spectrometry (IMMS) has gained popularity in the analysis of complex mixtures such as those encountered in metabolomics and proteomics. However, the challenge that exists in conventional pulsed IMMS is its inherent low duty cycle. The first application of Hadamard transform (HT)-type signal coupled with atmospheric pressure IMMS to complex mixtures is presented. Performance of the prototype was assessed by the analysis of metabolite standard mixture. With 200 times increased IMS duty cycle in HT mode compared with conventional pulsed mode, the limit of detection (LOD) was decreased by ∼10 times. Evaluation for application to complex mixtures was achieved using the NIST Standard Reference Material 1950 Metabolites in Human Plasma. Approximately 180 metabolite ions were detected within 1 min with an IMS resolving power (Rp) of ∼100. Rapid chromatographic separation prior to IMMS analysis was also demonstrated for improving the response of metabolite ions in rat brain tissue extract.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.