Abstract

Several similarities have been found between shear stress-induced erythrocyte damage and physiological aging of erythrocytes in terms of elevated mechanical fragility, increased erythrocyte aggregation, and decreased membrane surface charge. Accordingly, we hypothesized that blood pump circulation, which generates shear stress, would accelerate erythrocyte aging, manifesting as oxidation. Therefore, the purpose of this study was to investigate the effect of blood pump circulation on erythrocyte oxidation. Fresh porcine blood was acquired from a slaughterhouse and anticoagulated with sodium citrate. About 500 mL of anticoagulated whole blood was circulated for 180 min in an in vitro test circuit comprising a BP-80 blood pump with a pump speed and a pump pressure head of 100-120 mmHg. A blood sample was taken at the start of the circulation and 180 min afterward. The hemolysis level and oxidation amount of the erythrocyte membrane were analyzed and compared between samples. Hemolysis increased with the prolongation of shear exposure inside the pump circuit. After 180 min of blood pumping in circuit, the oxidation level of the erythrocyte membrane showed an increase of 0.1 nmol/mg protein. Moreover, the membrane oxidation levels of sheared erythrocytes were greater than those of control erythrocytes. These results suggest that blood pump circulation accelerates erythrocyte aging and give us a greater understanding of the effects of blood pump perfusion.