Abstract
A differential absorption lidar system for measurement of trace atmospheric substances in the visible and UV regions is developed. The system is based on a pair of Nd:YAG pumped dye lasers, each capable of emitting two wavelengths on alternate shots. The measurement error of the system was evaluated by measurement of tropospheric O3 using two identical differential absorption lidar (DIAL) pairs, from which two null profiles and four DIAL profiles are obtained simultaneously. The null profiles, obtained for both the on and off wavelengths used in DIAL measurement, are useful to evaluate systematic DIAL measurement error due to beam misalignment. The null profiles show that this systematic error is at most 1.3%. The total DIAL measurement error is about 5%, most of which can be attributed to statistical error.