Abstract
This paper describes a technique for the evaluation of complex residual errors in coaxial and waveguide vector network analyzer (VNA) measurements in the range of microwave, millimeter and submillimeter wave frequencies. The technique is based on reference values of scattering parameters for transmission line standards consisting of coaxial air dielectric lines and waveguide standard sections. Reference values traceable to SI-base units were calibrated for the standard transmission lines by the National Metrology Institute of Japan (NMIJ). Complex residual analyses are shown for VNA measurements with a PC-7 coaxial line system and with WR-15, WR-10 and WR-3 waveguide systems up to 330 GHz.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.