Abstract

Three commercially available test systems for the identification of nonfermentative gram-negative rods were compared: API 20NE, flow N/F and Minitek Nonfermenter System. Two hundred strains were identified by conventional means and by each test system. The rate of correct identification of Acinetobacter, Alcaligenes, Flavobacterium and Moraxella strains to genus level and of the other genera to species level was 92% with API 20NE, 84% with flow N/F and 75% with Minitek Nonfermenter System. The need for these kits in the diagnostic hospital laboratory is also discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.