Abstract
The Analytab Products Inc. (API), Micro-Drop (MD), and Uni-Yeast-Tek (UYT) systems for the presumptive identification of common clinical yeast isolates were compared with the oxidation-fermentation (OF) and a conventional procedure. With 229 coded isolates, the identification accuracies were API 94, MD 83, OF 82, and UYT 99%. The API system required the greatest technical ability. The MD materials were prone to malfunction. OF media, if incubated beyond 14 days, gave an accuracy of 87%, but this offered no advantage over the conventional procedure. The UYT system was the easiest to use.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.