Abstract

Modern radiotherapy treatment techniques commonly include multi-leaf collimators (MLCs) to shape the treatment fields and to conform the radiation dose to the target volume. MLCs require accurate and frequent quality assurance (QA) to ensure spatial and temporal accuracy of the leaves in order to allow optimal dose delivery to the patient. In this study, the accuracy and efficacy of AutoCAL, a commercial software for MLC QA, were evaluated. The software was found to be reproducible to within 0.2mm and to correspond with conventional QA methods (within 1.2 and 0.9mm of film and water tank measurements, respectively.) Thus, AutoCAL was found to be an accurate and efficient tool for routine MLC QA and calibration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.